Measurement of Comfort Levels by Means of Electrical Stapedial Reflex in Children
Mary Bresnihan, AFRCSI;
Gary Norman, MSc;
Fiona Scott, MSc;
Laura Viani, FRCSI, BSc
Arch Otolaryngol Head Neck Surg. 2001;127:963-966.
ABSTRACT
 |  |
Background Patient success and satisfaction with a cochlear implant largely depend
on the adequacy of the speech-processing program. The program is generated
by means of behaviorally determined threshold and comfort levels for each
electrode. As the minimum age for implantation continues to decrease, behavioral
methods of measuring comfort levels have become more problematic, and so the
need for objective ways to program speech processors has become more important.
Objectives To evaluate the use of electrically evoked stapedial reflexes (ESRs)
to measure comfort levels for children and compare these results with behavioral
measurements, and to report the results of a questionnaire assessing the acceptability
and general performance of program before and after adjustment of comfort
levels measured with ESRs.
Design and Setting Before-after trial in the cochlear implant unit of a tertiary hospital.
Patients and Methods Programming with the ESR technique was successfully completed in 20
of a consecutive sample of 26 children undergoing programming of their cochlear
implants.
Outcome Measures Programming units as measured by the 2 programming techniques and numerical
score of questionnaire.
Results Comfort levels with the ESR method were found to be consistently lower
than those obtained with behavioral techniques. Children using programs set
with ESRs wore their implants longer and had fewer episodes of discomfort
to environmental sounds.
Conclusion Comfort level estimation by means of ESRs is reliable and objective
and hence a valuable programming tool in the pediatric population.
INTRODUCTION
COCHLEAR implants require programming on an individual basis to provide
appropriate levels of electrical stimulation. This program, or "map," is placed
in the speech processor of the cochlear implant. Success of implantation largely
depends on the adequacy of this map. The map is generated by means of an electrical
dynamic range for each active electrode. This range is established by ascertaining
a threshold (T) measurement, which serves as the lower limit, and a comfort
(C) level, which is the upper limit of electrical stimulation. The T level
is defined as the lowest stimulus level producing 100% detection. The C level
is obtained by increasing the current from the T level until the patient indicates
that the stimulus is loud but comfortable. These levels vary from electrode
to electrode as well as among patients. They also vary over time and from
one programming session to the next.
Traditionally, the C and T levels are set by means of behavioral methods.
These methods are similar to those used in obtaining an audiogram and usually
require the child to pair a stimulus with a response. Setting the level of
minimal stimulation, the T level, is usually successfully achieved by standard
pediatric audiology techniques. These include operant conditioning tasks,
conditioned play audiometry, and visual reinforcement audiology. Children
who have undergone cochlear implantation are usually familiar with such techniques.
Although most young children can be conditioned to provide reliable T levels,
obtaining C levels from children can be extremely difficult. Setting the T
level requires the child to respond to the presence or absence of a sound.
Although the same techniques are used in setting the C level, the child is
now required to make a judgment about the sound beyond its simple presence
or absence. Young children often lack the attention span, patience, and cognitive
and language skills necessary to perform these tasks with good reliability
and repeatability. Moreover, the concept of "most comfortable level" is often
beyond young children's capabilities, even among the hearing population.
Implant clinicians (ie, audiologists, medical physicists, and physicians)
often rely on intuition and trial and error when deciding on a C level, and
the lack of auditory experience in very young children, together with limited
language abilities, makes the task of setting T and C levels a long-term,
ongoing process. Overstimulation or understimulation of the device can occur
easily and may lead to lack of appropriate progress. As the minimum age for
implantation has continued to decrease, successfully achieving the C level
is becoming more problematic. Clearly, the need for objective ways to program
speech processors has become more important. A reliable, objective method
would improve the accuracy and efficiency of mapping the speech processor.
The electrically evoked stapedial reflex (ESR) has been shown by researchers
to correlate with C levels in the adult population.1-4
The ability to elicit the acoustic reflex by means of electrical stimulation
in cochlear implant recipients has been established since 1986.5
Shallop and Ash6 showed that ESR correlated
with the most comfortable level rather than the uncomfortable level in an
adult population, and Spivak et al7 showed
ESR to be safely below the threshold for discomfort. Hodges et al8 demonstrated that maps generated by means of ESR to
measure C levels resulted in speech-perception performance that matched that
of behaviorally derived maps in an adult test group. Therefore, using ESR
as an objective predictor of C levels is potentially a very useful tool in
the pediatric population. Although measurement of ESR is clearly practical
and efficient, there has been little research into the outcome of patients
whose maps have been adjusted by means of ESR. Spivak et al7
looked at subjects' performance on 2 measures of speech recognition and found
that the maps based on ESR had acceptable sound quality and speech perception
and compared favorably with maps based on behavioral C levels. However, to
our knowledge, no researchers to date have looked at the outcome after altering
the map in the pediatric population.
The aims of this study were to compare the results of behaviorally vs
objectively recorded measurements of C levels with the use of ESR and to establish
whether there were any changes in the child's behavior or response to sound
when a program with C levels set by ESR was used.
PATIENTS AND METHODS
PATIENTS
The study included all children who were scheduled for programming in
August and September 1999. Verbal consent was obtained from all parents. A
total of 26 children were tested. The mean age was 4.9 years (range, 2-9 years),
and the male-female ratio was 15:11. The mean length of implant use was 18.5
months (range, 2-36 months). Twenty children were using the Nucleus 24 cochlear
implant with the Sprint processor (Cochlear Ltd, Sydney, Australia) and 6
were using the Nucleus 22 with the Spectra processor (Cochlear Ltd). All children
were using the Spectral Peak speech processing strategy (Cochlear Ltd).
PROCEDURE
A middle ear analyzer (GSI 33, version 2.0, model 1733; Grayson Stadler,
Milford, Ohio) was used to perform tympanometry and record the ESRs. All children
underwent tympanometry before ESR testing. The Nucleus 22 speech processors
were connected to the Nucleus diagnostic programming system, while the Nucleus
24 processors were connected to the clinical programming system. The recording
system of the middle ear analyzer is very sensitive to movement. To minimize
artifacts caused by movement, passive cooperation was obtained from the children.
They were seated in front of a color television and provided with an age-appropriate
video to watch, or they were allowed to read. Parents were encouraged to bring
a selection of their child's favorite cartoons or movies to the clinic.
After impedance checks and behavioral T-level measurements, each child
had a short break and his or her processor was reactivated with their current
map. Parents were encouraged during this interval to take the child for a
short walk. We believe that this reduces the large contrast between listening
at threshold levels and listening at suprathreshold measurements.
The recording of ESR was attempted in the ear contralateral to that
implanted. An appropriate probe tip was inserted in the ear; occasionally,
smearing the cuff with petroleum jelly provided a more stable and consistent
seal. The probe case was retained on the strap and placed over the child's
or parent's shoulder.
Once a normal tympanogram was obtained, the middle ear analyzer was
set to the "special (reflex decay)" mode and the system was pressurized. Stimulation
was derived via the appropriate interface by means of standard biphasic pulses
and presented through the child's own speech processor at the standard rate
of 250 pulses per second. A change in acoustic admittance for the 226-Hz probe's
tone resulting from the stapedial reflex contraction was presented as a downward
deflection of the prestimulus baseline.
Stimulation began at 20 programming units below the previously behaviorally
measured C level. The number of presentations ranged from 1 to 3. For the
Nucleus 24 users, this was controlled by the keyboard command for WinDPS (Windows
Diagnostic Programming System) (version R116.02 [build 445], service release
2; Cochlear Ltd). For Nucleus 22 users, control was through the diagnostic
programming system (version 6-10F; Cochlear Ltd). If a clear reflex was present,
the stimulus level was decreased by 5 units. If no reflex was evident, stimulation
was increased by 2 units until a reflex was present or there were behavioral
signs of the signal producing loudness discomfort. The ESR threshold was taken
as the lowest stimulus level that produced a deflection in baseline recording,
synchronous with the stimulus presentation.
The C level was set at 2 programming units below the level at which
the ESR threshold was identified. All available electrodes were tested.
STATISTICAL ANALYSIS
Statistical analysis used the paired means t
test with the use of Minitab for Windows, version 12.22 (Minitab Inc, State
College, Pa).
RESULTS
The ESR technique could not be performed in 6 patients (23%). Middle
ear effusion was present in the contralateral ear in 2 of these children.
Three children reported loudness discomfort before reaching ESR threshold;
all 3 of these had a history of previous middle ear disease. Sufficient cooperation
was not obtained in 1 child on 2 separate occasions to allow the procedure
to be performed. The 20 remaining children successfully completed ESR measurement.
A total of 391 individual measurements were made. Nine electrodes were not
tested because of nonauditory stimulation.
BEHAVIORALLY VS OBJECTIVELY RECORDED MEASUREMENTS OF C LEVELS BY ESR
The T levels were remeasured by behavioral methods to ensure that they
remained consistent with the T levels obtained in the previous programming
session. There was no statistical difference between the T level measurements
obtained during the 2 programming sessions (Figure 1).
|
|
|
|
Figure 1. Comparison between threshold levels.
BT1 and BT2 indicate threshold levels measured by behavioral testing on 2
occasions.
|
|
|
The 2 methods of measuring the C level were compared and the average
measurement for each electrode of the 20 children was calculated (Figure 2). The C level obtained with ESRs
was found to be consistently lower across all electrodes. Figure 3 shows the difference in C levels (behavioral - ESR);
a positive value indicates that the behavioral C level was higher than ESR
C level, whereas a negative value indicates that the ESR C level was higher.
The minimum and maximum levels represent the range. The difference in C levels
with the 2 measurements was found to be statistically significant across all
electrodes. Of the 391 measurements made, 380 were positive values and only
11 were negative values.
|
|
|
|
Figure 2. Comparison of 2 methods of comfort
level measurement. ESR indicates electrically evoked stapedial reflex.
|
|
|
|
|
|
|
Figure 3. Difference in comfort levels (subjective -
objective).
|
|
|
OUTCOME OF NEW PROGRAM WITH ESR C LEVEL
Parents were asked to complete a questionnaire relating to the new programming
technique compared with that used on all the previous visits, ie, the behavioral
method. Answers were completed in the form of a numerical analog scale ranging
from 0 to 4, where 0 represented "never" and 4 represented "always." Parents
were required to rate the original program set based on behavioral methods
and the new program set based on ESR. The questions consisted of 2 subgroups:
the first group of questions (Figure 4) asked whether the child ever showed discomfort to a range of 8 different environmental
sounds. The next 12 questions (Figure 5) pertained to the child's willingness to wear the implant and are outlined
in Table 1. Across all the questions,
the program based on ESR was rated better than behavioral programming, and
the differences for 10 of the questions were statistically significant (Figure 4 and Figure 5).
|
|
|
|
Figure 4. Parent responses to the first
set of questions rating the behavioral method vs the electrically evoked stapedial
reflex (ESR) method. A score of 0 indicates never; 4, always.
|
|
|
|
|
|
|
Figure 5. Parent responses to the second
set of questions rating the behavioral method vs the electrically evoked stapedial
reflex (ESR) method. A score of 0 indicates never; 4, always.
|
|
|
|
|
|
|
Questions Pertaining to Child Wearing Implant
|
|
|
COMMENT
Testing of ESR is not universally performed in programming cochlear
implants. It is objective and relatively straightforward to carry out. Therefore,
if it were to be accepted as a feasible method of programming, use of the
ESR should improve the ease of programming and, more important, the quality
of the speech-processing map. If measurement of ESR is to replace traditional
behavioral methods for setting C levels, certain criteria need to be met:
(1) ESR maps must not result in unacceptable levels of discomfort, (2) ESR
maps must result in the same or superior implant use, and (3) sound quality
should be similar or better than that produced by traditional behavioral techniques.
The T levels were assessed to ensure that the 2 programming sessions
were comparable despite being performed on different days. When C levels obtained
with ESR were compared with those obtained with behavioral methods, the ESR
levels were found to be consistently below behavioral levels. As a result,
we can be certain that the children are not being overstimulated through the
use of the ESR technique. Only 11 measurements across all electrodes in all
20 patients were above those established by behavioral techniques. In 1 patient
there was a difference of as much as 80 programming units, and this child
was taking his implant off regularly despite the fact that previously repeated
measurements obtained with behavioral techniques had shown consistent C levels.
Our findings are similar to those found in previous studies, although our
results were more consistent and had less variation.
As the new C levels were all set at a lower level, we were concerned
that the child's ability to use the implant may be compromised. The questionnaire
that was filled out by the parents demonstrated that the majority of children
found the new map superior to the original one, in particular, in situations
that would imply that the child was being overstimulated with the C levels
obtained by behavioral techniques. The degree of difference in the 2 programs
was surprising and was statistically significant in 50% of questions. Although
this was not a randomized controlled trial, the parents were unanimous in
their belief that programming with ESR results in improved usage and less
discomfort than traditional techniques.
CONCLUSIONS
Testing of ESRs is a clinically feasible procedure in patients with
no middle ear effusion and was successfully used in creating the map for the
speech processor. From the clinician's perspective, it is an effective procedure
that is easier to perform than behavioral tasks. It is both safe and reliable,
and one can be confident that the sound is audible but not uncomfortable.
From the patient's perspective, the new program results in fewer situations
that cause discomfort; therefore, patients are likely to wear the implant
more. Since this study, measuring ESR has replaced behavioral techniques in
programming the speech processor in our department and has become routine
procedure. We believe that it is a superior technique and should replace traditional,
subjective techniques when cochlear implants are programmed in the pediatric
population.
AUTHOR INFORMATION
Accepted for publication March 17, 2001.
Presented at the American Society of Pediatric Otolaryngology meeting,
Orlando, Fla, May 18, 2000.
Corresponding author and reprints: Mary Bresnihan, AFRCSI, Cochlear
Implant Unit, Beaumont Hospital, PO Box 1297, Dublin 9, Ireland (e-mail:
mary_bresnihan{at}hotmail.com).
From the Cochlear Implant Programme, Beaumont Hospital, Dublin, Ireland.
REFERENCES
 |  |
1. Battmer R-D, Laszig R, Lenhardt E. Electrically elicited stapedius reflex in cochlear implant patients. Ear Hear. 1990;11:370-374.
PUBMED
2. Hodges AV. The Relationship Between Electrically Evoked Responses
and Psychophysical Percepts Obtained Through a Nucleus 22 Channel Cochlear
Implant [dissertation]. Charlottesville: University of Virginia; 1990.
3. Jerger JF, Oliver TA, Chmiel RA. Prediction of dynamic range from stapedius reflex in cochlear implant
patients. Ear Hear. 1988;9:4-8.
PUBMED
4. Spivak LG, Chute PM. The relationship between electrical acoustic reflex thresholds and
behavioral comfort levels in children and adult cochlear implant patient. Ear Hear. 1994;15:184-192.
PUBMED
5. Jerger J, Jenkins H, Fifer R, Mecklenburg D. Stapedius reflex to electrical stimulation in a patient with a cohlear
implant. Ann Otol Rhinol Laryngol. 1986;95:151-157.
PUBMED
6. Shallop JK, Ash KR. Relationships among comfort levels determined by cochlear implant patient's
self programming, audiologist's programming, and electrical stapedius reflex
thresholds. Ann Otol Rhinol Laryngol. 1995;104(suppl 166, pt 2):175-176.
7. Spivak LG, Chute PM, Popp AL, Parisier SC. Programming the cochlear implant based on electrical acoustic reflex
thresholds: patient performance. Laryngoscope. 1994;104:1225-1230.
PUBMED
8. Hodges AV, Balkeny TJ, Ruth RA, Lambert PR, Dolan-Ash S, Schloffman JJ. Electrical middle ear muscle reflex: use in cochlear implant programming. Otolaryngol Head Neck Surg. 1997;117:255-261.
PUBMED
RELATED ARTICLE
Archives of OtolaryngologyHead & Neck Surgery Reader's Choice: Continuing Medical Education
Arch Otolaryngol Head Neck Surg. 2001;127(8):1011-1012.
FULL TEXT
|