You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT ARCHIVES
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 124 No. 3, March 1998 TABLE OF CONTENTS
  Archives
  •  Online Features
  Original Article
 This Article
 •Abstract
 •PDF
 • Reply to article
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citing articles on HighWire
 •Citing articles on Web of Science (7)
 •Contact me when this article is cited
 Related Content
 •Similar articles in this journal
 Topic Collections
 •Neurology
 •Neuro-otology
 •Hearing Loss/ Deafness
 •Alert me on articles by topic
 Social Bookmarking
  Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit Add to Technorati Add to Twitter What's this?

Short- and Long-term Results With Implantable Transcutaneous and Percutaneous Bone-Conduction Devices

Ad F. M. Snik, PhD; Wouter A. Dreschler, PhD; Rinze A. Tange, MD, PhD; Cor W. R. J. Cremers, MD, PhD

Arch Otolaryngol Head Neck Surg. 1998;124:265-268.

ABSTRACT

Objectives  To compare the percutaneous bone-anchored hearing aid (BAHA; type NBC-HC-200, Nobel Biocare, Gothenburg, Sweden) and the transcutaneous temporal bone stimulator (TBS; Xomed-Treace, Jacksonville, Fla) with conventional hearing aids and to evaluate long-term results.

Design  In a prospective clinical study, the new implantable bone-conduction devices were compared with the patients' previous conventional hearing aids. Speech perception in quiet and in noise were studied, and a questionnaire concerning the actual use of the device and speech recognition was administered. During follow-up that exceeded 4 years, relevant technical and medical problems were documented.

Patients  Forty-one successive subjects who were fitted with a BAHA and 17 subjects who were fitted with a TBS.

Results  In most subjects who had previously used a bone-conduction device, the new BAHA and TBS devices led to improved or comparable results on speech recognition tests and the questionnaire. However, among the subjects who had previously used air-conduction hearing aids, the results were ambiguous. In the long-term, the percentage of nonusers in the BAHA group was 5% (2/39); in the TBS group, 65% (13/20). The main reasons for not using the TBS were insufficient gain and medical and technical problems. The vulnerability of the percutaneous coupling of the BAHA to trauma or inflammation was not a major issue; only 4 implants were lost during the total follow-up of more than 250 years.

Conclusion  Results indicate that the BAHA is the better choice.



INTRODUCTION
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

SOME SUBJECTS who are hard of hearing and need a hearing aid can only be fitted with a bone-conduction hearing aid, because chronic draining of the ears or aural atresia make it impossible to fit hearing aids with ear molds. For many years, only 1 type of bone-conduction hearing aid was available, which transmitted the sound vibrations to the skull by means of a vibrator pressed on the skin in the mastoid region. In the early 1980s, Hakansson et al1-2 developed a bone-conduction hearing aid whose vibrator was anchored directly to the skull using a percutaneous titanium implant (BAHA; Nobel Biocare, Gothenburg, Sweden). Compared with the conventional bone-conduction device, the main advantages of the BAHA are that the transducer does not press against the skin, and that there are no sound-attenuating tissue layers between the vibrator and the skull. Hakansson et al1 reported that going through the skin and subcutaneous layers resulted in much more effective sound transmission, especially in the high-frequency range.

Hough et al3 developed a transcutaneous bone-conduction device, the temporal bone stimulator (TBS) or audiant bone-conductor (Xomed-Treace, Jacksonville, Fla). In their mechanism, a permanent magnet is implanted in the temporal bone and covered by a thin layer of skin. This magnet is driven by an external coil positioned on the skin and kept in place by a small coupling magnet. The main advantage of this method compared with the BAHA seems to be less vulnerability to trauma and infections (no skin-penetrating implant). The main disadvantage is the relatively wide distance between the implanted magnet and the driving coil, which results in loss of power.4

Although several studies have compared the TBS or the BAHA with conventional devices,2, 5-11 very few studies have included both new types of bone-conduction hearing aid. Wade et al12 presented the results in 24 subjects with a TBS and 11 subjects with a BAHA. They reported that on average, the free-field thresholds with the BAHA were better than those with the TBS. Several subjects with a TBS did not use the device regularly because of insufficient gain. Only 1 subject with a BAHA did not use it, according to the authors, because of "a phobia against wearing it." The authors suggested applying a TBS only in subjects with normal cochlear function, whereas the BAHA would be suitable for subjects with a sensorineural hearing loss component of up to 45 dB hearing level (HL).

We herein summarize our short- and long-term results with these direct bone-conduction devices. The BAHA was applied by our group at Nijmegen University Hospital, Nijmegen, the Netherlands, whereas the TBS was applied by our group at the Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands. Prospectively, strict protocols were developed under the supervision of the Dutch Council for Investigative Medicine of the Ziekenfondsraad, so that patient selection and the evaluation procedures were similar at both clinics. The selection criteria included binaural hearing loss of the conductive or mixed type with no surgical options for hearing improvement and rejection of a conventional bone-conduction hearing aid due to pain or skin irritation caused by the pressure of the bone-conduction vibrator.

According to the inclusion criteria of the manufacturer, the TBS was applied to subjects with a sensorineural hearing loss component of at most 25 to 30 dB HL. For the standard BAHA (type NBC-HC-200), the upper limit was 40 dB HL, but our study only included subjects with a sensorineural hearing loss component in the TBS application range.

The short-term evaluation compared the newly fitted bone-conduction device and the previous individually fitted conventional hearing aid. The subjects' performance on speech recognition tests and their opinions on speech recognition in quiet and noisy surroundings were included. Data concerning the previous device were obtained before surgery. Short-term performance was investigated relatively soon after fitting the new device within the 2-month trial, whereas long-term results concerned the actual use of the new device.


SUBJECTS, MATERIALS, AND METHODS
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

SUBJECTS

The results of all the patients who fulfilled the inclusion criteria and were fitted with the new bone-conduction device from January 1, 1989, to August 31, 1992, were included in the study. The study group consisted of 41 subjects who were fitted with a standard BAHA and 20 subjects who were fitted with a TBS. The TBS users were initially fitted with the more powerful, body-worn processor. If there was sufficient reserve gain, an ear-level processor was applied.

Three of the 20 TBS users did not undergo evaluation.

Before application of the new bone-conduction device, all subjects had used hearing aids. In the BAHA group, 33 subjects had used conventional bone-conduction hearing aids, and 8 had used air-conduction hearing aids. In the TBS group, these figures were 9 and 8, respectively. Some subject characteristics are presented in Table 1; the subjects in the BAHA and TBS groups were not matched. However, the mean values were comparable, and there was large overlap in the ranges.


View this table:
[in this window]
[in a new window]
Table 1. Subject Characteristics*


MATERIALS AND METHODS

Before and shortly after fitting the new bone-conduction device, speech perception tests were performed to compare the subjects' performance with their individually fitted BAHA or TBS and with the previous hearing aid. The new device was evaluated after a trial of at least 6 weeks. The test used was the Sentence Recognition in Noise Test, as described by Plomp and Mimpen.13 This test uses lists of 13 sentences and a steady-state, speech-shaped noise. The noise is presented at a fixed level, and the speech reception threshold of the sentences is established with an adaptive procedure. The difference between the speech reception threshold and the noise level (in decibels) is the critical speech-to-noise (S-N) ratio. Furthermore, the speech reception threshold was determined without noise, ie, speech reception threshold in quiet (SRT). In contrast to the SRT, the S-N ratio is independent of the volume setting of the hearing aid, as long as the speech level is above the subject's threshold.

The change in the S-N ratio following the change in devices can be expressed as a change in the percentage of correctly repeated sentences knowing that change of 1 dB in the S-N ratio equals 17% change in sentence recognition.13

We used a questionnaire to obtain subjective opinions about the previous and new hearing aids. It was administered twice, before surgery for the previous hearing aid and 3 to 5 months after fitting for the new bone-conduction device. The questionnaire has been described in detail elsewhere.14 We considered the questions on the recognition of speech in relatively quiet surroundings (5 subquestions) and in noisy situations (9 subquestions). The subjects answered the questions on a scale from 1 (impossible) to 10 (excellent). For each subject, the average score on the subquestions was calculated for both sets of items. During follow-up, the subjects were regularly asked about the actual use of the new device, and any relevant medical and technical problems were documented.


RESULTS
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

AFTER THE TRIAL PERIOD

Table 2 presents the change in SRT, namely the SRT with the previous device minus that with the new device, averaged for each subgroup. Among BAHA and TBS users who had previously used a conventional bone-conduction hearing aid, the mean SRT improved. The improvement was significant only in the BAHA users. Among subjects who had previously used an air-conduction hearing aid, the mean SRT deteriorated. This deterioration was significant in the BAHA group. The median change in the score for the question on speech recognition in quiet surroundings showed improvement in the BAHA subjects who had previously used bone-conduction hearing aids. In the other 3 subgroups, the median change was near zero.


View this table:
[in this window]
[in a new window]
Table 2. Main Outcome of Study*


The improvement in S-N ratio with the new device is also presented in Table 2. Whether the change in the S-N ratio was significantly different from zero is indicated. In the 2 BAHA subgroups, significant improvement was found, which was not the case in the TBS subgroups.

The median change in the score for the question on speech recognition in noisy surroundings is also presented. Among the BAHA users who had previously used bone-conduction devices, the median change was 1.6, which means that the BAHA was an obvious improvement over the previous conventional bone-conduction device. The median improvement in the BAHA users who had previously used air-conduction hearing aids was zero. In the TBS group, minor improvements were found.

Figure 1 presents the percentages of BAHA and TBS users who preferred their new bone-conduction device or their previous hearing aid or did not have any preference. Combining the results for the questions on speech recognition in quiet and in noisy surroundings, it can be concluded that in the BAHA users who had previously used bone-conduction hearing aids, there was obvious preference for the BAHA (about 70% preferred the BAHA while only 12% preferred the conventional bone-conduction hearing aid). About half of the BAHA users who had previously used air-conduction hearing aids preferred the BAHA, whereas somewhat less than half preferred the air-conduction hearing aid. Among the TBS users who had previously used a conventional bone conductor, about 50% preferred the TBS, and no one preferred the previous device. About 30% of the TBS users who had previously used air-conduction hearing aids preferred the previous device, while the others had no preference or preferred the TBS.



View larger version (28K):
[in this window]
[in a new window]
The percentage of users of the percutaneous titanium implant (BAHA; Nobel Biocare, Gothenburg, Sweden) and temporal bone stimulator (TBS; Xomed-Treace, Jacksonville, Fla) who, according to the short-term questionnaire, preferred the new bone-conduction device or their previous device or did not have any preference with regard to speech recognition in quiet (SIQ) or in noisy surroundings (SIN). The results of the 4 subgroups are presented separately. CBCHA indicates previous users of conventional bone-conduction hearing aid; ACHA, previous users of air-conducted hearing aid.


After the trial, all the BAHA users chose to use the BAHA, not their previous hearing aid. As mentioned above, 3 (15%) of the initial 20 subjects with a TBS did not undergo evaluation. These 3 patients experienced insufficient gain and returned to their previous devices within the trial before the first evaluation session. The sensorineural hearing loss component of these 3 patients was well within the manufacturer's inclusion criteria.

LONG-TERM RESULTS IN BAHA USERS

Two of the subjects in the BAHA group died of causes not related to their hearing problems. The long-term evaluation includes the remaining 39 subjects.

One subject had been using the BAHA successfully for more than 2 years, but then lost the implant because of inflammation. It was decided not to replace the implant because of psychological problems. Another subject had been using the BAHA for almost 3 years but experienced pain when using the device. No explanation could be found for the complaints, and at the subject's request, the implant was removed. One subject lost the implant because of inflammation and 2 subjects, because of trauma. In these 3 subjects, the implant was replaced and has been used without any further problems for 41/2, 51/2, and 7 years. In 2 other patients, it was decided to reduce further the thickness of the subcutaneous layer around the implant to minimize the risk for inflammation. Altogether, 6 reoperations were performed.

Rejection of the BAHA because of insufficient amplification was not encountered. However, 1 subject suffered severe deterioration in the sensorineural hearing loss component from 25 to more than 65 dB HL after surgery for cholesteatoma in the cerebellopontine angle. She was refitted with the more powerful BAHA type NBC-HC-220, but because of the severe deterioration of cochlear function, the result was poor.

At the last evaluation after at least 4 1/2 years of follow-up, 2 (5%) of the 39 BAHA recipients were nonusers. All the other subjects were using their BAHA on a daily basis.

LONG-TERM RESULTS IN TBS USERS

Medical problems with the TBS implant were encountered in 3 of the 17 subjects. The implanted magnet was removed from 1 patient because of inflammation and from 2 patients because of complaints about headaches and/or pressure on the skull. Beyond the trial period, 10 patients stopped using their device. Besides the medical problems mentioned above, lack of motivation and the vulnerability of the device that led to many repairs were reasons to discontinue use. In some of the older subjects, gradually amplification became insufficient because cochlear function deteriorated. Including the 3 patients who stopped using their TBS during the trial period, the total number of nonusers was 13 (65%) of 20 patients.


COMMENT
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

After the trial period, the results showed that in audiological terms, with some restrictions, application of both new types of bone-conduction device was successful. The results with the BAHA were somewhat better than those with the TBS. On average, it was found that in the subjects who had previously used a bone-conduction device, the new bone-conduction devices led to improved or comparable results on the tests as well as the questionnaire. However, among the subjects who had previously used air-conduction hearing aids, the results were ambiguous. This does not mean that an air-conduction hearing aid is still the first choice, because these subjects had been advised not to use their air-conduction hearing aid any longer owing to chronic draining of the ears. However, in such cases, before it is decided to change to a bone-conduction device, good counseling and a thorough audiological evaluation, including the use of a power CROS (contralateral routing of signals) air-conduction device with a vented ear-mold, are needed.

The long-term success rate with the BAHA was considerably higher than that with the TBS. Nonuse because of insufficient amplification was only encountered in the TBS group. In several other studies of TBS users who had sensorineural hearing loss within the manufacturer's criteria,3 similar results were found.6, 9, 12 The limited gain and output of the TBS, as shown by Gatehouse and Browning,15 may have played a role. The number of nonusers among the BAHA users was low, as in other studies.5, 7, 11-12 Rejection of the BAHA because of insufficient amplification was not encountered. In addition, in the case of deterioration of hearing loss, the more powerful BAHA type NBC-HC-220 can be fitted. This BAHA has been applied with success to subjects with a sensorineural hearing loss component of up to 65 dB HL.10

The percutaneous coupling of the BAHA might be a disadvantage, because it is vulnerable to trauma and inflammation. However, long-term clinical studies have shown that the titanium implant is reliable and stable in time.5, 11, 16 Nevertheless, implant loss may occur. In our study, with a total follow-up of more than 250 years, 4 implants were lost because of inflammation or trauma, which can be considered acceptable.

Encouraging long-term results were found with the BAHA, but not with the TBS. The good BAHA results are of great importance for subjects in whom the fitting of conventional hearing aids is troublesome because of aural atresia or chronic draining of the ears. The TBS should be applied only if the application of a percutaneous implant is contraindicated, eg, if regular adequate cleaning of the skin around the implant can not be guaranteed. Furthermore, to prevent rejection of the TBS because of insufficient gain, cochlear function should be normal or subnormal.


AUTHOR INFORMATION
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

Accepted for publication December 4, 1997.

Supported by grants from the Fund of Investigative Medicine of the Ziekenfondsraad of the Netherlands.

We thank Andy J. Beynon, MSc, Theo Zuidema, PhD, Emmanuel A. M. Mylanus, PhD, and Hanneke v.d. Vijsel for their assistance.

Reprints: Ad F. M. Snik, PhD, Department of Otorhinolaryngology, Nijmegen University Hospital, PO Box 9101, 6500 HB Nijmegen, the Netherlands (e-mail: a.snik{at}kno.azn.nl).

From the Departments of Otorhinolaryngology, Nijmegen University Hospital, Nijmegen (Drs Snik and Cremers), and Academic Medical Center, University of Amsterdam, Amsterdam (Drs Dreschler and Tange), the Netherlands.


REFERENCES
 Jump to Section
 •Top
 •Introduction
 •Subjects, materials, and methods
 •Results
 •Comment
 •Author information
 •References

1. Hakansson B, Tjellstrom A, Rosenhall U. Hearing thresholds with direct bone conduction versus conventional bone conduction. Scand Audiol. 1984;13:3-13. WEB OF SCIENCE | PUBMED
2. Hakansson B, Tjellstrom A, Rosenhall U, Carlsson P. The bone-anchored hearing aid: principles design and a psycho-acoustical evaluation. Acta Otolaryngol (Stockh). 1985;100:229-239. FULL TEXT | PUBMED
3. Hough JV, Vernon J, Johnson B, Dormer K, Himelick T. Experiences with implantable hearing devices and a presentation of a new device. Ann Otol Rhinol Laryngol. 1986;95(1, pt 1):60-65.
4. Hakansson B, Tjellstrom A, Carlsson P. Percutaneous versus transcutaneous transducers for hearing by direct bone conduction. Otolaryngol Head Neck Surg. 1990;102:339-344. WEB OF SCIENCE | PUBMED
5. Hakansson B, Liden B, Tjellstrom A, et al. Ten years of experience with the Swedish bone-anchored hearing system. Ann Otol Rhinol Laryngol. 1990;99(suppl 151):1-16.
6. Browning GG. The British experience of an implantable, subcutaneous bone-conduction hearing aid (Xomed Audiant). J Laryngol Otol. 1990;104:534-538. WEB OF SCIENCE | PUBMED
7. Bonding P, Jonsson MH, Salomon G, Algren P. The bone-anchored hearing aid: osseointegration and audiological effect. Acta Otolaryngol Suppl. 1992;492:42-45. PUBMED
8. Zuidema T, Van den Berg R, Dreschler WA, Tange RA. Implantable bone conductors: Dutch experiences with a transcutaneous device. Semin Hear. 1992;13:347-356.
9. van der Hulst RJAM, Dreschler WA, Tange RA. Experiences with an implantable bone-conduction hearing aid. Eur Arch Otorhinolaryngol. 1993;250:69-72. PUBMED
10. Snik AFM, Mylanus EAM, Cremers CWRJ. The bone-anchored hearing aid compared with conventional hearing aids: audiologic results and the patients' opinions. Otolaryngol Clin North Am. 1995;28:73-83. WEB OF SCIENCE | PUBMED
11. Cooper H, Burrell S, Powell R, Proops DW, Bickerton J. The Birmingham bone anchored hearing aid programme: referrals, selection, rehabilitation, philosophy and adult results. J Laryngol Otol Suppl. 1996;21:13-20. PUBMED
12. Wade PS, Halik JJ, Chasin M. Bone-conduction implants: transcutaneous versus percutaneous. Otolaryngol Head Neck Surg. 1992;106:68-74. WEB OF SCIENCE | PUBMED
13. Plomp R, Mimpen AM. Improving the reliability of testing the speech perception threshold for sentences. Audiology. 1979;18:43-52. WEB OF SCIENCE | PUBMED
14. Mylanus EAM, Snik AFM, Cremers CWRJ. Patient's opinions of bone anchored vs conventional hearing aids. Arch Otolaryngol Head Neck Surg. 1995;121:421-425. FREE FULL TEXT
15. Gatehouse S, Browning GG. The output characteristics of an implanted bone-conduction prosthesis. Clin Otolaryngol . 1990;15:503-513. FULL TEXT | WEB OF SCIENCE | PUBMED
16. Mylanus EAM, Cremers CW, Snik AF, van den Berge NW. Clinical results of percutaneous implants in the temporal bone. Arch Otolaryngol Head Neck Surg. 1994;120:81-85. FREE FULL TEXT


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter     What's this?

THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Electronic aids to hearing
Gatehouse
Br Med Bull 2002;63:147-156.
ABSTRACT | FULL TEXT  





HOME | CURRENT ISSUE | PAST ISSUES | TOPIC COLLECTIONS | CME | SUBMIT | SUBSCRIBE | HELP
CONDITIONS OF USE | PRIVACY POLICY | CONTACT US | SITE MAP
 
© 1998 American Medical Association. All Rights Reserved.