The Air Caloric TestA Parametric Study
Alfred C. Coats, MD;
Foster Hebert;
G. Robert Atwood
Arch Otolaryngol. 1976;102(6):343-354.
Abstract
 |  |
We studied the normal air caloric response's intensity and variability. Intensities of "preliminary standard" (47 C, 25 C, 60 seconds, 10 liters/min) air and standard bithermal water responses were the same, but the air responses were more variable. Control experiments suggested that the tendency of the air stream to equilibrate toward ambient may account for much of air's relatively high variability. Increasing the air irrigation's duration, flow rate, and separation from body temperature increased response intensity. Air's test-retest variability decreased when flow rate and duration increased and when temperature approached ambient. Therefore, we suggested that irrigating temperatures be chosen as close to ambient as possible and duration and flow rate be made as large as possible. From a table equating air and standard bithermal water responses, we selected tentative clinical air irrigation parameters of 27.5 C, 45.5 C, 100 seconds, and 13 liters/min.
(Arch Otolaryngol 102:343-354, 1976)
Author Affiliations
From the departments of physiology and otorhinolaryngology and communicative sciences, Baylor College of Medicine, Houston.
Footnotes
Accepted for publication Dec 29, 1975.
Reprint requests to the Department of Physiology, Baylor College of Medicine, 1200 Moursund, Houston, TX 77025 (Dr Coats).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Variability of Air Calorics vs Water: Statistical Implications
Cooper and Mason
Arch Otolaryngol Head Neck Surg 1979;105:113-115.
ABSTRACT
Reliabilities of Air and Water Caloric Responses
Ford and Stockwell
Arch Otolaryngol Head Neck Surg 1978;104:380-382.
ABSTRACT
|